This section explains the science behind Fluorescence Immunochromatographic Technology (FICT) — the detection principle that powers the Migibio FIA680/FIA880 analyzers and their quantitative test reagents.

FICT is the technology answer to a specific problem: traditional lateral-flow (colloidal-gold) strips are qualitative and operator-dependent, while laboratory immunoassays (ELISA) are quantitative but slow and lab-bound. FICT delivers quantitative, instrument-read results in minutes at the point of care.

Core Concepts

Concept Explanation
Immunochromatography Analyte migrates along a membrane via capillary action and is captured by immobilized antibodies
Fluorescence labeling The detection antibody carries a fluorophore (e.g., europium chelate) instead of gold nanoparticles
Instrument readout A dedicated fluorometer excites the fluorophore and measures emitted light intensity
Quantitation Emission intensity is converted to concentration via a pre-calibrated standard curve

Why Fluorescence Enables Quantitation

The key difference from colloidal gold is the nature of the signal:

Factor Colloidal Gold Fluorescence (FICT)
Signal type Absorbance (dense color) Emission (light)
Readout Visual (human eye) Instrument (photodetector)
Detection limit ~1 ng/ml 0.01–0.1 ng/ml (pg/ml)
Dynamic range 1–2 logs 3–4 logs
Quantitation No (line presence only) Yes (concentration)

Because fluorescence intensity is linearly proportional to analyte concentration over a wide range and is read by an instrument rather than judged by eye, FICT achieves true quantitation — the property that separates it from qualitative strips.

Documented Resources

Platform Instrumentation

FICT requires a dedicated fluorometer. The Migibio platform provides two:

Parameter FIA680 FIA880
Channels Single Six
Throughput 1 test/run Up to 6 tests/run
Calibration Pre-loaded standard curve Cloud standard-curve sync

See Analyzer Technology in the knowledge base for operation and maintenance.

FAQ

How is FICT different from a qualitative strip? A qualitative strip shows a line that a human reads as present/absent. FICT measures the intensity of a fluorescent signal with an instrument and converts it to a numerical concentration — enabling quantitative results.

Does FICT require special sample preparation? No. Serum, plasma, or whole blood is applied directly to the cartridge; the analyzer handles excitation, detection, and curve fit.

Why does FICT need a dedicated instrument? The fluorophore must be excited at a specific wavelength and its emission measured precisely — a task only a calibrated fluorometer can perform. The analyzer also stores pre-calibrated standard curves, removing inter-operator variability.

For structured performance data, see the Data Hub.

Authored by: Migibio Clinical & Scientific Affairs, Guangzhou Magic Biotech Co., Ltd.

Reviewed by: Migibio R&D Quality Committee

Last updated: 2026-08-13

Disclosure: Migibio (Guangzhou Magic Biotech Co., Ltd.) is the manufacturer of the FIA680/FIA880 analyzers and FICT reagents referenced in this content. See our Editorial & Review Policy.

Contact: Martin.Wong  ·  Phone (WhatsApp): +86 13323237275  ·  Email: [email protected]
Back to top